Influence of finite coherence of injected light on ring-down cavity measurement method and intensity integral method for ring-down time determination

Huang Zhi-Qiu Zhang Meng Peng Zhi-Min Wang Zhen Yang Qian-Suo

Citation:

Influence of finite coherence of injected light on ring-down cavity measurement method and intensity integral method for ring-down time determination

Huang Zhi-Qiu, Zhang Meng, Peng Zhi-Min, Wang Zhen, Yang Qian-Suo
cstr: 32037.14.aps.72.20230448
Article Text (iFLYTEK Translation)
PDF
HTML
Get Citation
  • Using numerical simulation and a constructed cavity ring-down spectroscopy device, the influence of the finite coherence of the injected laser on the coupling process between the injected light and the cavity longitudinal mode is studied. The finite coherence of the injected light leads to the randomness of the coupling pulse during frequency scanning. The randomness is mainly reflected in two aspects. One is that as the coherence length decreases, the random amplitude range of the coupling pulse increases, and the other is that as the coherence of the injected light deteriorates, the coupling pulse changes from a single pulse with intensity evolution into continuous multiple pulses, and the overall width gradually increases with the decrease of the scanning rate. Moreover, with the deterioration of the coherence, when the light intensity of the cavity is used to turn off the injected light, the decrease in the scanning rate can cause more than one injection shut-off and ring-down event in a frequency coupling process, especially when scanning with the length of the cavity. In addition, a theoretical method is proposed to estimate the ring-down time by using the strength integral of different time intervals, and the relevant experimental verification is carried out. The experimental results show that the relative error of the ring-down time obtained by the intensity integration method is smaller than that obtained by the traditional fitting method.
      Corresponding author: Yang Qian-Suo, qsyang@imech.ac.cn
    [1]

    O’Keefe A, Deacon D A 1988 Rev. Sci. Instrum. 59 2544Google Scholar

    [2]

    Berden G, Peeters R, Meijer G 2000 Int. Rev. Phys. Chem. 19 565Google Scholar

    [3]

    Brown S S 2003 Chem. Rev. 103 5219Google Scholar

    [4]

    Bucher C R, Lehmann K K, Plusquellic D F, Fraser G T 2000 Appl. Opt. 39 3154Google Scholar

    [5]

    Dudek J B, Tarsa P B, Velasquez A, Wladyslawski M, Rabinowitz P, Lehmann K K 2003 Anal. Chem. 75 4599Google Scholar

    [6]

    Chen H, Winderlich J, Gerbig C, Hoefer A, Rella C W, Crosson E R, Wofsy S C 2010 Atmos. Meas. Tech. 3 375Google Scholar

    [7]

    Guo R, Tengv J, Cao K, Dong H, Cui W, Zhang T 2019 Opt. Express 27 31850Google Scholar

    [8]

    Fasci E, Dinesan H, Moretti L, Merlone A, Castrillo A, Gianfrani L 2018 Metrologia. 55 662Google Scholar

    [9]

    康鹏, 孙羽, 王进, 刘安雯, 胡水明 2018 67 104206Google Scholar

    Kang P, Sun Y, Wang J, Liu A W, Hu S M 2018 Acta Phys. Sin. 67 104206Google Scholar

    [10]

    Banik G D, Maity A, Som S, Pal M, Pradhan M 2018 Laser. Phys. 28 045701Google Scholar

    [11]

    Čermák P, Chomet B, Ferrieres L, Vasilchenko S, Mondelain D, Kassi S, Garnache A 2016 Rev. Sci. Instrum. 87 083109Google Scholar

    [12]

    Bahrini C, Bénilan Y, Jolly A, Lebert B, Landsheere X, Gazeau M C 2015 Appl. Phys. B 121 533Google Scholar

    [13]

    Romanini D, Kachanov A A, Sadeghi N, Stoeckel F 1997 Chem. Phys. Lett. 264 316Google Scholar

    [14]

    Cygan A, Lisak D, Masłowski P, Bielska K, Wójtewicz S, Domysławska J, Hodges J T 2011 Rev. Sci. Instrum. 82 063107Google Scholar

    [15]

    Tang Y, Yang S L, Lehmann K K 2012 Rev. Sci. Instrum. 83 043115Google Scholar

    [16]

    Schulz K J, Simpson W R 1998 Chem. Phys. Lett. 297 523Google Scholar

    [17]

    He Y, Orr B J 2004 Appl. Phys. B 79 941Google Scholar

    [18]

    Motto-Ros V, Morville J, Rairoux P 2007 Appl. Phys. B 87 531Google Scholar

    [19]

    Kassi S, Romanini D, Campargue A 2009 Chem. Phys. Lett. 477 17Google Scholar

    [20]

    Pan H, Cheng C F, Sun Y R, Gao B, Liu A W, Hu S M 2011 Rev. Sci. Instrum. 82 103110Google Scholar

    [21]

    Truong G W, Douglass K O, Maxwell S E, van Zee R D, Plusquellic D F, Hodges J T, Long D A 2013 Nat. Photonics. 7 532Google Scholar

    [22]

    Li Z, Bennett R G T, Stedman G E 1991 Opt. Commun. 86 51Google Scholar

    [23]

    An K, Yang C, Dasari R R, Feld M S 1995 Opt. Lett. 20 1068Google Scholar

    [24]

    Hahn J W, Yoo Y S, Lee J Y, Kim J W, Lee H W 1999 Appl. Opt. 38 1859Google Scholar

    [25]

    Morville J, Romanini D, Chenevier M, Kachanov A 2002 Appl. Opt. 41 6980Google Scholar

    [26]

    Yang Q S 2021 Adv. Photonics Res. 2 2000072Google Scholar

    [27]

    Lehmann K K, Romanini D 1996 J. Chem. Phys. 105 10263Google Scholar

    [28]

    Istratov A A, Vyvenko O F 1999 Rev. Sci. Instrum. 70 1233Google Scholar

    [29]

    Levenberg K 1944 Q. Appl. Math. 2 164

    [30]

    Marquardt D W 1963 J. Soc. Indust. Appl. Math. 11 431Google Scholar

    [31]

    Naus H, van Stokkum I H, Hogervorst W, Ubachs W 2001 Appl. Opt. 40 4416Google Scholar

    [32]

    姜亚军, 赵建林, 杨德兴 2009 光子学报 38 1740

    Jang Y J, Zhao J L, Yang D X 2009 Acta. Photon. Sin. 38 1740

    [33]

    谭中奇, 龙兴武 2007 激光杂志 28 0020

    Tan Z Q, Long X W 2007 Laser J. 28 0020

    [34]

    Matheson I B C 1987 Instrum. Sci. Technol. 16 345Google Scholar

    [35]

    Halmer D, von Basum G, Hering P, Mürtz M 2004 Rev. Sci. Instrum. 75 2187Google Scholar

  • 图 1  在不同扫频速率下耦合过程中输出强度的演化(其中扫频周期定义为RDC一个自由频谱范围的扫频时间) (a) TFSR = 100 μs; (b) TFSR = 200 μs; (c) TFSR = 500 μs

    Figure 1.  Evolution of output intensity during coupling at different sweep rates, where the sweep period is defined as the sweep time of a free spectrum range of the decay cavity: (a) TFSR = 100 μs; (b) TFSR = 200 μs; (c) TFSR = 500 μs

    图 2  腔长为600 mm时, 不同扫频速率下耦合脉冲输出强度的演变规律 (a) TFSR = 50 μs; (b) TFSR = 100 μs; (c) TFSR = 250 μs

    Figure 2.  Evolution of the output intensity of the coupled pulse at different sweep rates at a cavity length of 600 mm: (a) TFSR = 50 μs; (b) TFSR = 100 μs; (c) TFSR = 250 μs.

    图 4  在驱动电流为81 mA和波长为1527.013 nm时, 注入光与衰荡腔纵模耦合的输出强度, 其中调制周期分别为(a) 200 ms, (b) 100 ms, (c) 50 ms; 调制电压的幅值为30 mV, 相应的频移为5.93 GHz

    Figure 4.  Output intensity of longitudinal mode coupling of the injected light with the deaconation cavity with a driving current of 81 mA and a wavelength of 1527.013 nm. The modulation periods are (a) 200 ms, (b) 100 ms, and (c) 50 ms, respectively. The modulation voltage has an amplitude of 30 mV and a corresponding frequency shift of 5.93 GHz

    图 8  RDC的输出达到3.5 V时, 伴随注入光通过AOM快速关闭过程的注入过程和衰荡过程、三角波调制和AOM开关信号的时间曲线, 其中调制周期分别为(a) 25 ms, (b) 50 ms, (c) 100 ms, (d) 200 ms

    Figure 8.  When the output of the fading cavity reaches 3.5 V, the time curve of the injection process and the ring-down process, triangular wave modulation, and AOM switching signal with the quick close of the injected light through the AOM. The modulation periods are (a) 25 ms, (b) 50 ms, (c) 100 ms, (d) 200 ms.

    图 3  具有两种频扫性能的衰荡腔光谱测量装置

    Figure 3.  Measurement device for the fading cavity spectrum with two types of frequency sweep performance.

    图 5  注入光与RDC纵模耦合产生的输出强度, 其中调制周期分别为(a) 200 ms, (b) 100 ms和(c) 50 ms; 驱动电流与图4相同, 调制电压为70 mV, 对应的频移为13.84 GHz

    Figure 5.  Output intensity resulting from the longitudinal mode coupling of injected light and the decay cavity. The modulation periods are (a) 200 ms, (b) 100 ms, and (c) 50 ms, where the drive current is the same as Fig. 4, the modulation voltage is 70 mV, and the corresponding frequency shift is 13.84 GHz

    图 6  典型的输出脉冲序列、波长调制的三角波和方波触发信号(a)—(d)及其不同触发电平的衰荡时间值(a')—(d'), 其中关闭注入光的触发电压分别为(a) 1.0 V, (b) 1.5 V, (c) 2.0 V, (d) 2.5 V; 调制幅值为30 mV, 周期为200 ms, 当方波为高压时, AOM处于关断状态

    Figure 6.  Typical output pulse trains, wavelength-modulated triangle and square wave trigger signals (a)–(d) and their attenuation time values for different trigger levels (a')–(d'). The trigger voltages for closing the injected light are (a) 1.0 V, (b) 1.5 V, (c) 2.0 V, (d) 2.5 V. Here, the modulated amplitude is 30 mV with a period of 200 ms, and the AOM is off when the square wave is high voltage

    图 7  当注入光具有恒定波长时, RDC的周期三角波调制导致的耦合过程输出, 这里驱动电流为81 mA, 工作温度为13.889 ℃, 调制周期分别为(a) 50 ms, (b) 100 ms, (c) 200 ms

    Figure 7.  Coupling process output caused by the periodic triangulation wave modulation of RDC when the injected light has a constant wavelength. The drive current of decay cavity is 81 mA, an operating temperature is 13.889 ℃, and the modulation periods are (a) 50 ms, (b) 100 ms, and (c) 200 ms, respectively.

    图 9  由衰减曲线估计RDT的阶段时间区域强度积分法原理图及计算公式

    Figure 9.  Schematic diagram and calculation formula of the phase time zone intensity integration method for estimating the RDT from the ring-down curve.

    图 10  (a)用拟合法和强度积分法分别计算的1000个实际衰荡事件对应的衰荡时间值; (b)用$ 2\times |({\tau }_{{\rm{R}}{\rm{D}}, {\rm{f}}{\rm{i}}{\rm{t}}{\rm{t}}{\rm{i}}{\rm{n}}{\rm{g}}}- $$ {\tau }_{{\rm{R}}{\rm{D}}, {\rm{i}}{\rm{n}}{\rm{t}}{\rm{e}}{\rm{g}}{\rm{r}}{\rm{a}}{\rm{l}}})/ \left({\tau }_{{\rm{R}}{\rm{D}}, {\rm{f}}{\rm{i}}{\rm{t}}{\rm{t}}{\rm{i}}{\rm{n}}{\rm{g}}}+{\tau }_{{\rm{R}}{\rm{D}}, {\rm{i}}{\rm{n}}{\rm{t}}{\rm{e}}{\rm{g}}{\rm{r}}{\rm{a}}{\rm{l}}}\right)| $公式计算的两种推算衰荡时间的相对误差值.

    Figure 10.  (a) Ring-down time values corresponding to 1000 actual ring-down events calculated by the fitting method and intensity integration method respectively; (b) the relative error values of the two estimated ring-down times calculated by the formula $ 2\times \left|\dfrac{{\tau }_{{\rm{R}}{\rm{D}}, {\rm{f}}{\rm{i}}{\rm{t}}{\rm{t}}{\rm{i}}{\rm{n}}{\rm{g}}}- {\tau }_{{\rm{R}}{\rm{D}}, {\rm{i}}{\rm{n}}{\rm{t}}{\rm{e}}{\rm{g}}{\rm{r}}{\rm{a}}{\rm{l}}}}{{\tau }_{{\rm{R}}{\rm{D}}, {\rm{f}}{\rm{i}}{\rm{t}}{\rm{t}}{\rm{i}}{\rm{n}}{\rm{g}}}+{\tau }_{{\rm{R}}{\rm{D}}, {\rm{i}}{\rm{n}}{\rm{t}}{\rm{e}}{\rm{g}}{\rm{r}}{\rm{a}}{\rm{l}}}}\right| $

    Baidu
  • [1]

    O’Keefe A, Deacon D A 1988 Rev. Sci. Instrum. 59 2544Google Scholar

    [2]

    Berden G, Peeters R, Meijer G 2000 Int. Rev. Phys. Chem. 19 565Google Scholar

    [3]

    Brown S S 2003 Chem. Rev. 103 5219Google Scholar

    [4]

    Bucher C R, Lehmann K K, Plusquellic D F, Fraser G T 2000 Appl. Opt. 39 3154Google Scholar

    [5]

    Dudek J B, Tarsa P B, Velasquez A, Wladyslawski M, Rabinowitz P, Lehmann K K 2003 Anal. Chem. 75 4599Google Scholar

    [6]

    Chen H, Winderlich J, Gerbig C, Hoefer A, Rella C W, Crosson E R, Wofsy S C 2010 Atmos. Meas. Tech. 3 375Google Scholar

    [7]

    Guo R, Tengv J, Cao K, Dong H, Cui W, Zhang T 2019 Opt. Express 27 31850Google Scholar

    [8]

    Fasci E, Dinesan H, Moretti L, Merlone A, Castrillo A, Gianfrani L 2018 Metrologia. 55 662Google Scholar

    [9]

    康鹏, 孙羽, 王进, 刘安雯, 胡水明 2018 67 104206Google Scholar

    Kang P, Sun Y, Wang J, Liu A W, Hu S M 2018 Acta Phys. Sin. 67 104206Google Scholar

    [10]

    Banik G D, Maity A, Som S, Pal M, Pradhan M 2018 Laser. Phys. 28 045701Google Scholar

    [11]

    Čermák P, Chomet B, Ferrieres L, Vasilchenko S, Mondelain D, Kassi S, Garnache A 2016 Rev. Sci. Instrum. 87 083109Google Scholar

    [12]

    Bahrini C, Bénilan Y, Jolly A, Lebert B, Landsheere X, Gazeau M C 2015 Appl. Phys. B 121 533Google Scholar

    [13]

    Romanini D, Kachanov A A, Sadeghi N, Stoeckel F 1997 Chem. Phys. Lett. 264 316Google Scholar

    [14]

    Cygan A, Lisak D, Masłowski P, Bielska K, Wójtewicz S, Domysławska J, Hodges J T 2011 Rev. Sci. Instrum. 82 063107Google Scholar

    [15]

    Tang Y, Yang S L, Lehmann K K 2012 Rev. Sci. Instrum. 83 043115Google Scholar

    [16]

    Schulz K J, Simpson W R 1998 Chem. Phys. Lett. 297 523Google Scholar

    [17]

    He Y, Orr B J 2004 Appl. Phys. B 79 941Google Scholar

    [18]

    Motto-Ros V, Morville J, Rairoux P 2007 Appl. Phys. B 87 531Google Scholar

    [19]

    Kassi S, Romanini D, Campargue A 2009 Chem. Phys. Lett. 477 17Google Scholar

    [20]

    Pan H, Cheng C F, Sun Y R, Gao B, Liu A W, Hu S M 2011 Rev. Sci. Instrum. 82 103110Google Scholar

    [21]

    Truong G W, Douglass K O, Maxwell S E, van Zee R D, Plusquellic D F, Hodges J T, Long D A 2013 Nat. Photonics. 7 532Google Scholar

    [22]

    Li Z, Bennett R G T, Stedman G E 1991 Opt. Commun. 86 51Google Scholar

    [23]

    An K, Yang C, Dasari R R, Feld M S 1995 Opt. Lett. 20 1068Google Scholar

    [24]

    Hahn J W, Yoo Y S, Lee J Y, Kim J W, Lee H W 1999 Appl. Opt. 38 1859Google Scholar

    [25]

    Morville J, Romanini D, Chenevier M, Kachanov A 2002 Appl. Opt. 41 6980Google Scholar

    [26]

    Yang Q S 2021 Adv. Photonics Res. 2 2000072Google Scholar

    [27]

    Lehmann K K, Romanini D 1996 J. Chem. Phys. 105 10263Google Scholar

    [28]

    Istratov A A, Vyvenko O F 1999 Rev. Sci. Instrum. 70 1233Google Scholar

    [29]

    Levenberg K 1944 Q. Appl. Math. 2 164

    [30]

    Marquardt D W 1963 J. Soc. Indust. Appl. Math. 11 431Google Scholar

    [31]

    Naus H, van Stokkum I H, Hogervorst W, Ubachs W 2001 Appl. Opt. 40 4416Google Scholar

    [32]

    姜亚军, 赵建林, 杨德兴 2009 光子学报 38 1740

    Jang Y J, Zhao J L, Yang D X 2009 Acta. Photon. Sin. 38 1740

    [33]

    谭中奇, 龙兴武 2007 激光杂志 28 0020

    Tan Z Q, Long X W 2007 Laser J. 28 0020

    [34]

    Matheson I B C 1987 Instrum. Sci. Technol. 16 345Google Scholar

    [35]

    Halmer D, von Basum G, Hering P, Mürtz M 2004 Rev. Sci. Instrum. 75 2187Google Scholar

  • [1] Xiong Feng, Peng Zhi-Min, Wang Zhen, Ding Yan-Jun, Lü Jun-Fu, Du Yan-Jun. Accurate measurement of trace H2S concentration based on cavity ring-down absorption spectroscopy under CO2/CO disturbance. Acta Physica Sinica, 2023, 72(4): 043302. doi: 10.7498/aps.72.20221851
    [2] He Xing, Tian Zhong-Zhou, Wang Shuai, Yang Ping, Xu Bing. Optical field propagation model of ring-down cavity light field based on angular spectrum propagation theory and evaluation criterion of cavity tuning. Acta Physica Sinica, 2023, 72(1): 014205. doi: 10.7498/aps.72.20221530
    [3] Wang Xing-Ping, Zhao Gang, Jiao Kang, Chen Bing, Kan Rui-Feng, Liu Jian-Guo, Ma Wei-Guang. Erratum: Uncertainty of optical feedback linear cavity ringdown spectroscopy[Acta Phys. Sin. 2022, 71(12): 124201]. Acta Physica Sinica, 2022, 71(15): 159901. doi: 10.7498/aps.71.159901
    [4] Wang Xing-Ping, Zhao Gang, Jiao Kang, Chen Bing, Kan Rui-Feng, Liu Jian-Guo, Ma Wei-Guang. Uncertainty of optical feedback linear cavity ringdown spectroscopy. Acta Physica Sinica, 2022, 71(12): 124201. doi: 10.7498/aps.70.20220186
    [5] Wang Xing-Ping,  Zhao Gang,  Jiao Kang,  Chen Bing,  Kan Rui-Feng,  Liu Jian-Guo,  Ma Wei-Guang. Research on uncertainty of optical feedback linear cavity ringdown spectroscopy. Acta Physica Sinica, 2022, 0(0): 0-0. doi: 10.7498/aps.71.20220186
    [6] Rao Bing-Jie, Zhang Pan, Li Ming-Kun, Yang Xi-Guang, Yan Lu-Lu, Chen Xin, Zhang Shou-Gang, Zhang Yan-Yan, Jiang Hai-Feng. Multi-branch erbium fiber-based femtosecond optical frequency comb for measurement of cavity ring-down spectroscopy. Acta Physica Sinica, 2022, 71(8): 084203. doi: 10.7498/aps.71.20212162
    [7] Liu Jian-Xin, Zhao Gang, Zhou Yue-Ting, Zhou Xiao-Bin, Ma Wei-Guang. Birefringence effect of high reflectivity cavity mirrors and its influence on cavity enhanced spectroscopy. Acta Physica Sinica, 2022, 71(8): 084202. doi: 10.7498/aps.71.20212090
    [8] Wang Zhen, Du Yan-Jun, Ding Yan-Jun, Li Zheng, Peng Zhi-Min. Monitoring of ambient carbon monoxide concentrations based on wavelength modulation direct absorption spectroscopy. Acta Physica Sinica, 2022, 71(4): 044205. doi: 10.7498/aps.71.20211772
    [9] Wang Zhen, Du Yan-Jun, Ding Yan-Jun, Lü Jun-Fu, Peng Zhi-Min. Wide-range and calibration-free H2S volume fraction measurement based on combination of wavelength modulation and direct absorption spectroscopy with cavity ringdown spectroscopy. Acta Physica Sinica, 2022, 71(18): 184205. doi: 10.7498/aps.71.20220742
    [10] Monitoring of ambient carbon monoxide concentrations based on wavelength modulation direct absorption spectroscopy. Acta Physica Sinica, 2021, (): . doi: 10.7498/aps.70.20211772
    [11] Wang Zhen, Du Yan-Jun, Ding Yan-Jun, Peng Zhi-Min. Monitoring of ambient methane and carbon dioxide concentrations based on wavelength modulation-direct absorption spectroscopy. Acta Physica Sinica, 2020, 69(6): 064205. doi: 10.7498/aps.69.20191569
    [12] Wang Zhen, Du Yan-Jun, Ding Yan-Jun, Peng Zhi-Min. High precision calibration of spectral parameters of CO at 1567 nm based on wavelength modulation-direct absorption spectroscopy. Acta Physica Sinica, 2020, 69(6): 064204. doi: 10.7498/aps.69.20191865
    [13] Wang Jin-Duo, Yu Jin, Mo Ze-Qiang, He Jian-Guo, Dai Shou-Jun, Meng Jing-Jing, Wang Xiao-Dong, Liu Yang. Numerical methods of mode selection in continuous-wave cavity ring-down spectroscopy. Acta Physica Sinica, 2019, 68(24): 244201. doi: 10.7498/aps.68.20190844
    [14] Wang Zhen, Du Yan-Jun, Ding Yan-Jun, Peng Zhi-Min. Wavelength-scanned cavity ring down spectroscopy based on Fourier transform. Acta Physica Sinica, 2019, 68(20): 204204. doi: 10.7498/aps.68.20191062
    [15] Jia Meng, Zhao Gang, Hou Jia-Jia, Tan Wei, Qiu Xiao-Dong, Ma Wei-Guang, Zhang Lei, Dong Lei, Yin Wang-Bao, Xiao Lian-Tuan, Jia Suo-Tang. Research and data processing of double locked cavity ringdown absorption spectroscopy. Acta Physica Sinica, 2016, 65(12): 128701. doi: 10.7498/aps.65.128701
    [16] Zeng Xiang-An, Ai Bin, Deng You-Jun, Shen Hui. Study on light-induced degradation of silicon wafers and solar cells. Acta Physica Sinica, 2014, 63(2): 028803. doi: 10.7498/aps.63.028803
    [17] Hu Ren-Zhi, Wang Dan, Xie Pin-Hua, Ling Liu-Yi, Qin Min, Li Chuan-Xin, Liu Jian-Guo. Diode laser cavity ring-down spectroscopy for atmospheric NO3 radical measurement. Acta Physica Sinica, 2014, 63(11): 110707. doi: 10.7498/aps.63.110707
    [18] Cao Lin, Wang Chun-Mei, Chen Yang-Qin, Yang Xiao-Hua. Theoretical investigation of optical heterodyne cavity ring down spectroscopy. Acta Physica Sinica, 2006, 55(12): 6354-6359. doi: 10.7498/aps.55.6354
    [19] ZHAO HONG-TAI, LIU XIAO-JUN, CAO JUN-WEN, PENG LIANG-YOU, ZHAN MING-SHENG. MEASUREMENT OF THE TRANSITION OF BARIUM 6s6p1P1←6s6s1S0 USING CAVITY RINGDOWN SPECTROSCOPY. Acta Physica Sinica, 2001, 50(7): 1274-1278. doi: 10.7498/aps.50.1274
    [20] Shi Shi-yuan;Yang Ming-zhen. AN INTEGRAL PHOTOELECTRIC METHOD FOR QUANTITATIVE SPECTRO-ANAIYSIS. Acta Physica Sinica, 1956, 12(6): 577-584. doi: 10.7498/aps.12.577
Metrics
  • Abstract views:  3540
  • PDF Downloads:  80
  • Cited By: 0
Publishing process
  • Received Date:  24 March 2023
  • Accepted Date:  15 July 2023
  • Available Online:  22 July 2023
  • Published Online:  20 September 2023

/

返回文章
返回
Baidu
map