In cavity ring-down technique, cavity maladjustment has an essential effect on the measurement of intracavity loss. Several adjustment criterions have been adopted to achieve the optimal cavity state. However, experimental study shows that these criterions may correspond to different cavity states, which means that there is discrepancy between different criterions. In view of this problem, a model of intracavity propagation of Gaussian beam is established based on the angular spectrum propagation theory. This model is tested by numerical simulation and experimental research together. In the simulation, the true value of intracavity loss can be known beforehand. The two-dimensional angular scanning is carried out for certain cavity mirror. The two-dimensional distributions of the measure value of intracavity loss and the transmission light intensity are obtained simultaneously. These distributions are both nonlinear and multi-extremum, which will doubtlessly increase the difficulty in realizing the cavity adjustment. By comparing the distributions , we do find the discrepancy between the largest transmission light intensity and the least measured intracavity loss. Meanwhile both of these two states may be not corresponding to the true value in fact. After statistical studies, the relative error of the least measured intracavity loss is (–37.01±11.79) ppm, whereas the relative error of the largest transmission intensity is (–2.70±0.89) ppm. The criterion of the largest transmission intensity shows better stability and repeatability. This model is further tested in a folded cavity ring-down setup. The similar scanning procedure is carried out. A major problem in the experiment is that the true value of intracavity loss cannot be known. So only the repeatability precision of the measured intracavity loss can be analyzed. The statistical results of the largest light intensity and the least measured intracavity loss are ±29.32 ppm and ±70.71 ppm, respectively. The criterion of the largest transmission intensity has better repeatability, which is basically consistent with the simulation result. In this way the rationality of this model can be verified to some degree. In this paper, the criterion of the largest transmission intensity is recommended in the cavity ring-down technique. Furthermore, this model can be a reference for the research of intracavity optical field response, intracavity optical field transmission, unstable resonator alignment, etc.
-
Keywords:
- angular spectrum propagation /
- ring-down cavity /
- intracavity light field transmission /
- cavity maladjustment
[1] Sanders V 1977 Appl. Opt. 16 19Google Scholar
[2] 李斌成, 龚元 2010 激光与光电子学进展 47 021203Google Scholar
Li B C, Gong Y 2010 Laser Opt. Pro. 47 021203Google Scholar
[3] Tan Y, Wang J, Zhao X Q, Liu A W, Hu S M 2017 J. Quant. Spectrosc. Radiat. Transf. 187 274Google Scholar
[4] 康鹏, 孙羽, 王进, 刘安雯, 胡水明 2018 67 104206Google Scholar
Kang P, Sun Y, Wang J, Liu A W, Hu S M 2018 Acta Phys. Sin. 67 104206Google Scholar
[5] McHale L E, Hecobian A, Yalin A P 2016 Opt. Express 24 5523Google Scholar
[6] Li Z Y, Hu R Z, Xie P H, Chen H, Wu S Y, Wang F Y, Wang Y H, Ling L Y, Liu J G, Liu W Q 2018 Opt. Express 26 A433Google Scholar
[7] Yang L Z, Yang J J, Yang Y, Zhang Z W, Wang J F, Zhang Z X, Xue P P, Gong Y K, Copner N 2017 Opt. Express 25 2031Google Scholar
[8] Smith I W 1978 Appl. Opt. 17 2476Google Scholar
[9] Li B C, Gong Y 2010 US Patent 7 679 750 B2 [2010-03-16]
[10] Cui H, Li B C, Han Y L, Wang J, Gao C M, Wang Y F 2017 Chin. Opt. Lett. 15 053101Google Scholar
[11] Xiao S L, Li B C, Wang J 2020 Metrologia 57 055002Google Scholar
[12] Xiang W D, Yang P, Wang S, Xu B, Liu H 2018 Opto-Electronic Advances 1 180024Google Scholar
[13] Cui H, Li B C, Han Y L, Wang J, Gao C M, Wang Y F 2016 Opt. Express 24 013343Google Scholar
[14] Hamzeh T, Anam C P, Liu J J 2020 Appl. Opt. 59 9464Google Scholar
[15] Anderson D Z, Frisch J C, Masser C S 1984 Appl. Opt. 23 1238Google Scholar
[16] Paldus B A, Kachanov A A 2005 Cana. J. Phys. 83 975Google Scholar
[17] Romanini D 2014 Appl. Phys. B 115 517Google Scholar
[18] Shadman S, Rose C, Yalin A P 2016 Appl. Phys. B 122 194Google Scholar
[19] Maity A, Maithani S, Pradhan M 2021 Anal. Chem. 93 388Google Scholar
[20] 易亨瑜 2006 中国激光 33 399Google Scholar
Yi H Y 2006 Chin. J. Lasers 33 399Google Scholar
[21] 易亨瑜, 吕百达, 张凯 2006 激光技术 30 5Google Scholar
Yi H Y, Lv B D, Zhang K 2006 Laser Tech. 30 5Google Scholar
[22] 何星 2016 博士学位论文 (北京: 中国科学院大学)
He X 2016 Ph. D. Dissertation (Beijing: University of Chinese Academy of Sciences) (in Chinese)
[23] He X, Tian Z Z, Lai B H, Zhao W, Wang S, Yang P 2021 CN Patent 113984670A [2022-01-28]
[24] 薛颖, 杜星湖, 何星, 王帅, 杨平, 许冰 2020 中国激光 47 0504001Google Scholar
Xue Y, Du X H, He X, Wang S, Yang P, Xu B 2020 Chin. J. Lasers 47 0504001Google Scholar
[25] 杜星湖, 薛颖, 何星, 王帅, 杨平, 许冰 2020 中国激光 47 0604006Google Scholar
Du X H, Xue Y, He X, Wang S, Yang P, Xu B 2020 Chin. J. Lasers 47 0604006Google Scholar
[26] Hodges J T, Looney J P, Zee R D V 1996 J. Chem. Phys. 105 10278Google Scholar
[27] Lehmann K K 1996 J. Chem. Phys. 105 10263Google Scholar
[28] 吕乃光 2006 傅里叶光学(第二版) (北京: 机械工业出版社) 第82—86页
Lv N G 2006 Fourier Optics (Vol. 2) (Beijing: China Machine Press) pp82–86 (in Chinese)
[29] 曲哲超, 李斌成, 韩艳玲 2011 光子学报 40 1366Google Scholar
Qu Z C, Li B C, Han Y L 2011 Acta Phot. Sin. 40 1366Google Scholar
[30] He X, Yan H, Dong L Z, Yang P, Xu B 2016 Chin. Phys. B 25 014211Google Scholar
[31] He X, Luan Y S, Dong L Z, Yang P, Xu B, Tang G M 2016 Opto-Electronic Engineering 43 46Google Scholar
-
表 1 衰荡腔模型仿真初始参数列表
Table 1. The initial parameters of simulation model
名称 参数 数值 注入
光束横模模式 TEM00 中心波长 λ = 1064 nm 光源线宽* LW = 0.5 nm 束腰半径 w = 0.8 mm 衰荡腔 腔镜反射率 R1 = R2 = 99.85% 腔镜曲率 r1 = r2=1 m 腔长 L = 0.5 m 腔镜偏移量 $ \delta_{x1} =\delta_{71} =\delta_{x2} =\delta_{y2} = 0 $ 传输
过程注入过程 腔内往返传输2000次 衰荡过程 腔内往返传输500次 注: *为在光源线宽范围内, 离散选择30个波长进行仿真. 表 2 重复性扫描仿真结果
Table 2. Simulation results of repetitive scanning.
编号 腔镜M1的倾斜失调角度/(°) 光强最大峰值下的腔损耗/ppm 最长衰荡时间下的腔损耗/ppm 腔损耗真值/ppm θx1/(°) θy1/(°) 1 0 0.001 1498.06 1471.98 1500 2 –0.0005 0.001 1497.60 1461.02 1500 3 0.001 0.001 1497.51 1447.07 1500 4 0 0 1496.01 1471.87 1500 表 3 不同腔镜反射率下的腔损耗仿真结果
Table 3. Simulation results under different cavity mirror reflectivity.
腔镜反射率(R1=R2) 光强最大峰值下的腔损耗测量结果/ppm 最长衰荡时间下的腔损耗测量结果/ppm 99.80% (2000 ppm) 1996.66±0.80 1959.06±13.03 99.85% (1500 ppm) 1497.30±0.89 1462.99±11.79 99.88% (1200 ppm) 1197.63±1.06 1164.80±9.55 表 4 腔镜M1不同倾斜失调量下的实验结果
Table 4. Experimental results of cavity mirror M1 under different tilt maladjustment.
编号 光强最大峰值下
的腔损耗/ppm最长衰荡时间下
的腔损耗/ppm1 2240.61 2038.26 2 2261.66 2045.96 3 2304.96 1892.29 4 2245.25 1989.59 -
[1] Sanders V 1977 Appl. Opt. 16 19Google Scholar
[2] 李斌成, 龚元 2010 激光与光电子学进展 47 021203Google Scholar
Li B C, Gong Y 2010 Laser Opt. Pro. 47 021203Google Scholar
[3] Tan Y, Wang J, Zhao X Q, Liu A W, Hu S M 2017 J. Quant. Spectrosc. Radiat. Transf. 187 274Google Scholar
[4] 康鹏, 孙羽, 王进, 刘安雯, 胡水明 2018 67 104206Google Scholar
Kang P, Sun Y, Wang J, Liu A W, Hu S M 2018 Acta Phys. Sin. 67 104206Google Scholar
[5] McHale L E, Hecobian A, Yalin A P 2016 Opt. Express 24 5523Google Scholar
[6] Li Z Y, Hu R Z, Xie P H, Chen H, Wu S Y, Wang F Y, Wang Y H, Ling L Y, Liu J G, Liu W Q 2018 Opt. Express 26 A433Google Scholar
[7] Yang L Z, Yang J J, Yang Y, Zhang Z W, Wang J F, Zhang Z X, Xue P P, Gong Y K, Copner N 2017 Opt. Express 25 2031Google Scholar
[8] Smith I W 1978 Appl. Opt. 17 2476Google Scholar
[9] Li B C, Gong Y 2010 US Patent 7 679 750 B2 [2010-03-16]
[10] Cui H, Li B C, Han Y L, Wang J, Gao C M, Wang Y F 2017 Chin. Opt. Lett. 15 053101Google Scholar
[11] Xiao S L, Li B C, Wang J 2020 Metrologia 57 055002Google Scholar
[12] Xiang W D, Yang P, Wang S, Xu B, Liu H 2018 Opto-Electronic Advances 1 180024Google Scholar
[13] Cui H, Li B C, Han Y L, Wang J, Gao C M, Wang Y F 2016 Opt. Express 24 013343Google Scholar
[14] Hamzeh T, Anam C P, Liu J J 2020 Appl. Opt. 59 9464Google Scholar
[15] Anderson D Z, Frisch J C, Masser C S 1984 Appl. Opt. 23 1238Google Scholar
[16] Paldus B A, Kachanov A A 2005 Cana. J. Phys. 83 975Google Scholar
[17] Romanini D 2014 Appl. Phys. B 115 517Google Scholar
[18] Shadman S, Rose C, Yalin A P 2016 Appl. Phys. B 122 194Google Scholar
[19] Maity A, Maithani S, Pradhan M 2021 Anal. Chem. 93 388Google Scholar
[20] 易亨瑜 2006 中国激光 33 399Google Scholar
Yi H Y 2006 Chin. J. Lasers 33 399Google Scholar
[21] 易亨瑜, 吕百达, 张凯 2006 激光技术 30 5Google Scholar
Yi H Y, Lv B D, Zhang K 2006 Laser Tech. 30 5Google Scholar
[22] 何星 2016 博士学位论文 (北京: 中国科学院大学)
He X 2016 Ph. D. Dissertation (Beijing: University of Chinese Academy of Sciences) (in Chinese)
[23] He X, Tian Z Z, Lai B H, Zhao W, Wang S, Yang P 2021 CN Patent 113984670A [2022-01-28]
[24] 薛颖, 杜星湖, 何星, 王帅, 杨平, 许冰 2020 中国激光 47 0504001Google Scholar
Xue Y, Du X H, He X, Wang S, Yang P, Xu B 2020 Chin. J. Lasers 47 0504001Google Scholar
[25] 杜星湖, 薛颖, 何星, 王帅, 杨平, 许冰 2020 中国激光 47 0604006Google Scholar
Du X H, Xue Y, He X, Wang S, Yang P, Xu B 2020 Chin. J. Lasers 47 0604006Google Scholar
[26] Hodges J T, Looney J P, Zee R D V 1996 J. Chem. Phys. 105 10278Google Scholar
[27] Lehmann K K 1996 J. Chem. Phys. 105 10263Google Scholar
[28] 吕乃光 2006 傅里叶光学(第二版) (北京: 机械工业出版社) 第82—86页
Lv N G 2006 Fourier Optics (Vol. 2) (Beijing: China Machine Press) pp82–86 (in Chinese)
[29] 曲哲超, 李斌成, 韩艳玲 2011 光子学报 40 1366Google Scholar
Qu Z C, Li B C, Han Y L 2011 Acta Phot. Sin. 40 1366Google Scholar
[30] He X, Yan H, Dong L Z, Yang P, Xu B 2016 Chin. Phys. B 25 014211Google Scholar
[31] He X, Luan Y S, Dong L Z, Yang P, Xu B, Tang G M 2016 Opto-Electronic Engineering 43 46Google Scholar
Catalog
Metrics
- Abstract views: 3330
- PDF Downloads: 62
- Cited By: 0