Multifocal structured illumination microscopy (MSIM) can achieve optically sectioned images with twice the diffraction limited resolution at an imaging speed of 1 Hz and an imaging depth of up to 50 μm. Compared with the traditional wide-field SIM, the MSIM has greater imaging depth and optical sectionning ability, and it is more suitable for long-term three-dimensional (3D) super-resolution imaging of living thick samples. However, the MSIM has some problems, such as slow imaging speed and complex image post-processing process. In this work, a fast super-resolution imaging method and system based on the flat-field multiplexed MSIM (FM-MSIM) is proposed. By inserting a beam shaping device into the illumination light path, the Gaussian beam is reshaped into a uniform flat-top profile, thereby improving the intensity uniformity of excitation multi-spot focal array and expanding the field of view. By elongating each diffraction limited excitation focal point four times along the Y direction to form a new multiplexed multifocal array pattern, the number of scanning steps is reduced, the energy utilization is improved, and then the imaging speed and signal-to-noise ratio are improved. Combined with the sparse Bayesian learning image reconstruction algorithm based on multiple measurement vector model, the image reconstruction steps are simplified, the imaging speed can be improved at least 4 times while ensuring the spatial resolution of MSIM. On this basis, the established FM-MSIM system is used to carry out the super-resolution imaging experiments on the BSC cell microtubule samples and mouse kidney slices. The experimental results prove the fast three-dimensional super-resolution imaging ability of the system, which is of great significance in developing the fast MSIM.
-
Keywords:
- multifocal structured illumination microscopy /
- super-resolution imaging /
- flat-field illumination /
- Bayesian learning algorithm
[1] Hell S W, Wichmann J 1994 Opt. Lett. 19 780Google Scholar
[2] Klar T A, Jakobs S, Dyba M, Egner A, Hell S W 2000 Proc. Natl. Acad. Sci. U. S. A. 97 8206Google Scholar
[3] Willig K I, Harke B, Medda R, Hell S W 2007 Nat. Methods 4 915Google Scholar
[4] Rust M J, Bates M, Zhuang X 2006 Nat. Methods 3 793Google Scholar
[5] Betzig E, Patterson G H, Sougrat R, Lindwasser O W, Olenych S, Bonifacino J S, Davidson M W, Lippincott-Schwartz J, Hess H F 2006 Science 313 1642Google Scholar
[6] Gustafsson M G L 2000 J. Microsc. 198 82Google Scholar
[7] Gustafsson M G L 2005 Proc. Natl. Acad. Sci. U. S. A. 102 13081Google Scholar
[8] Gustafsson M G L, Shao L, Carlton P M, Wang C J, Golubovskaya I N, Cande W Z, Agard D A, Sedat J W 2008 Biophys. J. 94 4957Google Scholar
[9] Huang X, Fan J, Li L, Liu H, Wu R, Wu Y, Wei L, Mao H, Lal A, Xi P, Tang L, Zhang Y, Liu Y, Tan S, Chen L 2018 Nat. Biotechnol. 36 451Google Scholar
[10] Guo Y, Li D, Zhang S, Yang Y, Liu J J, Wang X, Liu C, Milkie D E, Moore R P, Tulu U S, Kiehart D P, Hu J, Lippincott-Schwartz J, Betzig E, Li D 2018 Cell 175 1430Google Scholar
[11] Muller C B, Enderlein J 2010 Phys. Rev. Lett. 104 198101Google Scholar
[12] York A G, Parekh S H, Dalle Nogare D, Fischer R S, Temprine K, Mione M, Chitnis A B, Combs C A, Shroff H 2012 Nat. Methods 9 749Google Scholar
[13] Schulz O, Pieper C, Clever M, Pfaff J, Ruhlandt A, Kehlenbach R H, Wouters F S, Grosshans J, Bunt G, Enderlein J 2013 Proc. Natl. Acad. Sci. U. S. A. 110 21000Google Scholar
[14] York A G, Chandris P, Nogare D D, Head J, Wawrzusin P, Fischer R S, Chitnis A, Shroff H 2013 Nat. Methods 10 1122Google Scholar
[15] De Luca G M, Breedijk R M, Brandt R A, Zeelenberg C H, de Jong B E, Timmermans W, Azar L N, Hoebe R A, Stallinga S, Manders E M 2013 Biomed. Opt. Express 4 2644Google Scholar
[16] Roth S, Heintzmann R 2016 Methods Appl. Fluoresc. 4 045005Google Scholar
[17] Wu J J, Li S W, Cao H Q, Lin D Y, Yu B, Qu J L 2018 Opt. Express 26 31430Google Scholar
-
图 2 平场照明实验表征 (a)无光束整形器的罗丹明均匀溶液宽场成像; (b)有光束整形器的罗丹明均匀溶液宽场成像; (c)图(a)和图(b)中蓝色虚线部分布归一化强度轮廓图; (d)无光束整形器的激发点阵; (e)有光束整形器的激发点阵
Figure 2. Experimental characterization of flat-field illumination: (a) Wide field imaging of uniform Rhodamine 6 G solution without a beam shaper; (b) wide field imaging of uniform Rhodamine 6 G solution with a beam shaper; (c) normalized intensity fitting profiles along blue dotted line in panel (a) and panel (b); (d) a multifocal excitation pattern without a beam shaper; (e) a multifocal excitation pattern with a beam shaper.
图 3 复用多焦点照明原理图 (a) 4 × 1激发点阵扫描原理图; (b) 1 × 1激发点阵扫描原理图; (c)复用的多焦点激发罗丹明均匀染料样品荧光图像
Figure 3. Schematic diagram of multiplexed multifocal excitation illumination: (a) Schematic diagram of 4 × 1 excitation spot array scanning; (b) schematic diagram of 1 × 1 excitation spot array scanning; (c) fluorescece image of the excitation foci in a uniform solution of Rhodamine 6G at the sample plane.
图 5 FM-MSIM 分辨率标定 (a)减去背景噪声的宽场微管图像; (b) 1 × 1点扫模板的MSBL重构图像; (c) 4 × 1点扫模板的MSBL重构图像; (d)蓝色实线处微管强度轮廓线; (e) 白色实线处微管强度轮廓线
Figure 5. Resolution in FM-MSIM: (a) Wide field image of microtubule in BSCs labeled with Alexa Fluor 488 phalloidin by subtracting background noise; (b) MSBL reconstructed image of 1 × 1 point scan mode; (c) MSBL reconstructed image of 4 × 1 point scan mode; (d) plots of intensity along blue solid line in panels (a)–(c); (e) plots of intensity along white solid line in panels (a)–(c).
图 6 小鼠肾切片不同轴向位置(z = 0, 3 , 6 μm)处的图像: (a)—(c)宽场图像; (d)—(f)MSBL重构图像; (g)图(a)和图(d)中蓝色虚线处的归一化强度分布轮廓曲线; (h) 图(b)和图(e)相同位置处归一化强度分布轮廓曲线; (i)图(c)和图(f)相同位置处归一化强度分布轮廓曲线
Figure 6. Images of mouse kidney section at different axial positions(z = 0, 3 and 6 μm): (a)–(c) Wide field images; (d)–(f) MSBL reconstruction images; (g) normalized intensity distribution profiles along the blue dotted line in panel (a) and panel (d); (h) normalized intensity distribution profiles along the same position in panel (b) and panel (e); (i) normalized intensity distribution profiles along the same position in panel (c) and panel (f).
-
[1] Hell S W, Wichmann J 1994 Opt. Lett. 19 780Google Scholar
[2] Klar T A, Jakobs S, Dyba M, Egner A, Hell S W 2000 Proc. Natl. Acad. Sci. U. S. A. 97 8206Google Scholar
[3] Willig K I, Harke B, Medda R, Hell S W 2007 Nat. Methods 4 915Google Scholar
[4] Rust M J, Bates M, Zhuang X 2006 Nat. Methods 3 793Google Scholar
[5] Betzig E, Patterson G H, Sougrat R, Lindwasser O W, Olenych S, Bonifacino J S, Davidson M W, Lippincott-Schwartz J, Hess H F 2006 Science 313 1642Google Scholar
[6] Gustafsson M G L 2000 J. Microsc. 198 82Google Scholar
[7] Gustafsson M G L 2005 Proc. Natl. Acad. Sci. U. S. A. 102 13081Google Scholar
[8] Gustafsson M G L, Shao L, Carlton P M, Wang C J, Golubovskaya I N, Cande W Z, Agard D A, Sedat J W 2008 Biophys. J. 94 4957Google Scholar
[9] Huang X, Fan J, Li L, Liu H, Wu R, Wu Y, Wei L, Mao H, Lal A, Xi P, Tang L, Zhang Y, Liu Y, Tan S, Chen L 2018 Nat. Biotechnol. 36 451Google Scholar
[10] Guo Y, Li D, Zhang S, Yang Y, Liu J J, Wang X, Liu C, Milkie D E, Moore R P, Tulu U S, Kiehart D P, Hu J, Lippincott-Schwartz J, Betzig E, Li D 2018 Cell 175 1430Google Scholar
[11] Muller C B, Enderlein J 2010 Phys. Rev. Lett. 104 198101Google Scholar
[12] York A G, Parekh S H, Dalle Nogare D, Fischer R S, Temprine K, Mione M, Chitnis A B, Combs C A, Shroff H 2012 Nat. Methods 9 749Google Scholar
[13] Schulz O, Pieper C, Clever M, Pfaff J, Ruhlandt A, Kehlenbach R H, Wouters F S, Grosshans J, Bunt G, Enderlein J 2013 Proc. Natl. Acad. Sci. U. S. A. 110 21000Google Scholar
[14] York A G, Chandris P, Nogare D D, Head J, Wawrzusin P, Fischer R S, Chitnis A, Shroff H 2013 Nat. Methods 10 1122Google Scholar
[15] De Luca G M, Breedijk R M, Brandt R A, Zeelenberg C H, de Jong B E, Timmermans W, Azar L N, Hoebe R A, Stallinga S, Manders E M 2013 Biomed. Opt. Express 4 2644Google Scholar
[16] Roth S, Heintzmann R 2016 Methods Appl. Fluoresc. 4 045005Google Scholar
[17] Wu J J, Li S W, Cao H Q, Lin D Y, Yu B, Qu J L 2018 Opt. Express 26 31430Google Scholar
Catalog
Metrics
- Abstract views: 5833
- PDF Downloads: 140
- Cited By: 0