-
With the advantages of non-ionizing and low cost, ultrasound imaging has been widely used in clinical diagnosis and treatment. However, due to the significant velocity changes between cortical bone and soft-tissue, the traditional ultrasound beamforming method under the assumption of constant velocity fails to reconstruct the cortical bone image. The velocity model based beamforming has been used in geophysics and non-destructive testing as an effective way to solve the challenges resulting from the velocity changes in multi-layer structure. Since the cortical bone can be modeled as a three-layer structure consisting of soft tissue, cortical bone and marrow, a multi-layer velocity model based synthetic aperture ultrasound method is introduced for cortical bone imaging. In this study, we first utilize synthetic transmit aperture ultrasound to obtain the full-matrix dataset to increase the signal-to-noise ratio. Second, a three-layer cortical bone velocity model is built with the compressed sensing estimated arriving time delay. The bases of compressed sensing consist of a series of excitation pulses with different delays. The received signals are regarded as a composition of the bases with different weights, thus can be projected into the bases by using compressed sensing. The time-delay of each received element is estimated by compressed sensing. According to the time-delay, the full-matrix dataset is reformed into a zero-offset format. By extracting the bases corresponded with the interface reflected signals, the time-delay between and the thickness values of the interfaces can be estimated. The velocity model can thus be built with the estimated cortical bone thickness. Based on the velocity model and zero-offset data, the phase shift migration method is used to reconstruct the cortical bone image. The finite-difference time-domain (FDTD) method is used to simulate the wave propagation in a 3.4-mm-thick cortical bone. The transmitting pulse is a Gaussian-function enveloped tone-burst signal with 6.25 MHz center frequency and 250 MHz iteration rate. The reconstructed image of simulation shows a clear top interface and bottom interface of cortical bone with correct thickness. Further FDTD simulations are carried out on a 3-mm-to-5-mm-thick cortical bone, and the average relative error of estimated thickness is 4.9% with a 13.5% variance. In vitro experiment is performed on a 3.4-mm-thick bovine bone plate to test the feasibility of the proposed method by using Verasonics platform (128-element linear array). The transmitting pulse is a Gaussian-function enveloped tone-burst signal with 6.25 MHz center frequency and 25 MHz sampling rate. The reconstructed image in experiment reveals a clear top interface and bottom interface of cortical bone with correct thickness. The experiment is repeated several times and the average relative error of estimated thickness is 3.6% with a 5.4% variance. The results of simulation and experiment both indicate that compressed sensing is effective in estimating the delay parameters of the velocity model. Finally, we evaluate the capability of compressed sensing in time-delay estimation, and the result shows that compressed sensing is more accurate than Hilbert transform even in a 20 dB-noise condition. In conclusion, the proposed method can be useful in the thickness estimation and the ultrasound imaging of cortical bone. In vivo experiment and clinical application should be further investigated.
-
Keywords:
- synthetic aperture ultrasound /
- cortical bone imaging /
- velocity model /
- compressed sensing
[1] Bover J, Bailone L, Lopez-Baez V, Benito S, Ciceri P, Galassi A, Cozzolino M 2017 J. Nephrol. 30 677Google Scholar
[2] 章轶立, 魏戌, 申浩, 谢雁鸣 2018 中国骨质疏松杂志 24 676Google Scholar
Zhang Y L, Wei X, Shen H, Xie Y M 2018 Chin. J. Osteoporos. 24 676Google Scholar
[3] Oo W M, Naganathan V, Bo M T, Hunter D J 2018 Quant. Imaging. Med. Surg. 8 100Google Scholar
[4] 他得安, 王威琪, 汪源源 2009 应用声学 28 161Google Scholar
Ta D A, Wang W Q, Wang Y Y 2009 Appl. Acoust. 28 161Google Scholar
[5] Minonzio J G, Bochud N, Vallet Q, Ramiandrisoa D, Etcheto A, Briot K, Kolta S, Roux C, Laugier P 2019 J. Bone Miner. Res.Google Scholar
[6] Ta D A, Wang W Q, Wang Y Y, Le L H, Zhou Y Q 2009 Ultrasound. Med. Biol. 35 641Google Scholar
[7] Xu K L, Minonzio J G, Ta D A, Hu B, Wang W Q, Laugier P 2016 IEEE Trans. Ultrason. Ferroelectr. Freq. Control 63 1514Google Scholar
[8] Xu K L, Ta D A, He R X, Qin Y X, Wang W Q 2014 Ultrasound. Med. Biol. 40 817Google Scholar
[9] Bai L, Xu K L, Li D, Ta D A, Le L H, Wang W Q 2018 J. Biomech. 77 83Google Scholar
[10] Xu K L, Laugier P, Minonzio J G 2018 J. Acoust. Soc. Am. 143 2729Google Scholar
[11] Chartier L B, Bosco L, Lapointe-Shaw L, Chenkin J 2017 CJEM 19 131Google Scholar
[12] 刘洋, 郭霞生, 章东, 龚秀芬 2011 声学学报 36 179Google Scholar
Liu Y, Guo X S, Zhang D, Gong X F 2011 Acta Acustica 36 179Google Scholar
[13] Li H J, Le L H, Sacchi M D, Lou E H M 2013 Ultrasound. Med. Biol. 39 1482Google Scholar
[14] Renaud G, Kruizinga P, Cassereau D, Laugier P 2018 Phys. Med. Biol. 63 125010Google Scholar
[15] Jensen J A, Nikolov S I, Gammelmark K L, Pedersen M H 2006 Ultrasonics 44 e5Google Scholar
[16] Yu M Y, Li Y, Ma T, Shung K K, Zhou Q F 2017 IEEE Trans. Med. Imaging 36 2171Google Scholar
[17] Nayak R, Schifitto G, Doyley M M 2017 Med. Phys. 44 4068Google Scholar
[18] Brandt A H, Hemmsen M C, Hansen P M, Madsen S S, Krohn P S, Lange T, Hansen K L, Jensen J A, Nielsen M B 2015 Ultrasound. Med. Biol. 41 2368Google Scholar
[19] Taner M T, Koehler F 1969 Geophysics 34 821
[20] Saenger E H, Kocur G K, Jud R, Torrilhon M 2011 Appl. Math. Modell. 35 807Google Scholar
[21] Gazdag J 1978 Geophysics 43 1342Google Scholar
[22] Olofsson T 2010 IEEE Trans. Ultrason. Ferroelectr. Freq. Control 57 2522Google Scholar
[23] 孙宝申, 沈建中 1993 应用声学 3 43Google Scholar
Sun B S, Shen J Z 1993 Appl. Acoust. 3 43Google Scholar
[24] Trots I, Nowicki A, Lewandowski M 2010 WASET 4 136
[25] 康荣宗, 田鹏武, 于宏毅 2014 63 200701Google Scholar
Kang R Z, Tian P W, Yu H Y 2014 Acta Phys. Sin. 63 200701Google Scholar
[26] 刘珍黎, 宋亮华, 白亮, 许凯亮, 他得安 2017 66 154303Google Scholar
Liu Z L, Song L H, Bai L, Xu K L, Ta D A 2017 Acta Phys. Sin. 66 154303Google Scholar
[27] Liu C C 2014 Ph. D. Dissertation (Shanghai: Fudan University) (in Chinese)
[28] Baron C, Talmant M, Laugier P 2007 J. Acoust. Soc. Am. 122 1810Google Scholar
[29] Pithioux M, Lasaygues P, Chabrand P 2002 J. Biomech. 35 961Google Scholar
[30] Qin K H, Yang C, Sun F 2014 IEEE Trans. Ultrason. Ferroelectr. Freq. Control 61 133Google Scholar
-
图 5 仿真结果 (a)压缩感知前30个基底; (b)单通道接收信号; (c)单次发射全部通道接收信号; (d)压缩感知调整的单次发射全部通道接收信号
Figure 5. Simulated results: (a) The first 30 bases of compressed sensing; (b) received signal of single element; (c) received signals of all elements; (d) compressed sensing based temporally adjusted received signals of all elements.
图 7 实验结果 (a)压缩感知前30个基底; (b)单通道接收信号; (c)单次发射全部通道接收信号; (d)压缩感知调整的单次发射全部通道接收信号
Figure 7. Experiment results: (a) The first 30 bases of compressed sensing; (b) received signal of single element; (c) received signals of all elements; (d) compressed sensing based temporally adjusted received signals of all elements.
表 1 皮质骨厚度估计及误差
Table 1. Estimation and relative error of cortical bone thickness.
参量 数值 真实厚度/mm 3.0 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4.0 4.1 4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9 估计厚度/mm 2.7 3.2 3.3 3.0 3.4 3.9 3.9 3.3 3.5 4.0 4.1 4.2 4.2 4.0 4.5 4.5 4.3 4.6 4.6 4.7 相对误差/% 10 3.2 3.1 9.1 0 11 8.3 11 7.9 2.6 2.5 2.4 0 6.9 2.3 0 6.5 2.1 4.2 4.1 -
[1] Bover J, Bailone L, Lopez-Baez V, Benito S, Ciceri P, Galassi A, Cozzolino M 2017 J. Nephrol. 30 677Google Scholar
[2] 章轶立, 魏戌, 申浩, 谢雁鸣 2018 中国骨质疏松杂志 24 676Google Scholar
Zhang Y L, Wei X, Shen H, Xie Y M 2018 Chin. J. Osteoporos. 24 676Google Scholar
[3] Oo W M, Naganathan V, Bo M T, Hunter D J 2018 Quant. Imaging. Med. Surg. 8 100Google Scholar
[4] 他得安, 王威琪, 汪源源 2009 应用声学 28 161Google Scholar
Ta D A, Wang W Q, Wang Y Y 2009 Appl. Acoust. 28 161Google Scholar
[5] Minonzio J G, Bochud N, Vallet Q, Ramiandrisoa D, Etcheto A, Briot K, Kolta S, Roux C, Laugier P 2019 J. Bone Miner. Res.Google Scholar
[6] Ta D A, Wang W Q, Wang Y Y, Le L H, Zhou Y Q 2009 Ultrasound. Med. Biol. 35 641Google Scholar
[7] Xu K L, Minonzio J G, Ta D A, Hu B, Wang W Q, Laugier P 2016 IEEE Trans. Ultrason. Ferroelectr. Freq. Control 63 1514Google Scholar
[8] Xu K L, Ta D A, He R X, Qin Y X, Wang W Q 2014 Ultrasound. Med. Biol. 40 817Google Scholar
[9] Bai L, Xu K L, Li D, Ta D A, Le L H, Wang W Q 2018 J. Biomech. 77 83Google Scholar
[10] Xu K L, Laugier P, Minonzio J G 2018 J. Acoust. Soc. Am. 143 2729Google Scholar
[11] Chartier L B, Bosco L, Lapointe-Shaw L, Chenkin J 2017 CJEM 19 131Google Scholar
[12] 刘洋, 郭霞生, 章东, 龚秀芬 2011 声学学报 36 179Google Scholar
Liu Y, Guo X S, Zhang D, Gong X F 2011 Acta Acustica 36 179Google Scholar
[13] Li H J, Le L H, Sacchi M D, Lou E H M 2013 Ultrasound. Med. Biol. 39 1482Google Scholar
[14] Renaud G, Kruizinga P, Cassereau D, Laugier P 2018 Phys. Med. Biol. 63 125010Google Scholar
[15] Jensen J A, Nikolov S I, Gammelmark K L, Pedersen M H 2006 Ultrasonics 44 e5Google Scholar
[16] Yu M Y, Li Y, Ma T, Shung K K, Zhou Q F 2017 IEEE Trans. Med. Imaging 36 2171Google Scholar
[17] Nayak R, Schifitto G, Doyley M M 2017 Med. Phys. 44 4068Google Scholar
[18] Brandt A H, Hemmsen M C, Hansen P M, Madsen S S, Krohn P S, Lange T, Hansen K L, Jensen J A, Nielsen M B 2015 Ultrasound. Med. Biol. 41 2368Google Scholar
[19] Taner M T, Koehler F 1969 Geophysics 34 821
[20] Saenger E H, Kocur G K, Jud R, Torrilhon M 2011 Appl. Math. Modell. 35 807Google Scholar
[21] Gazdag J 1978 Geophysics 43 1342Google Scholar
[22] Olofsson T 2010 IEEE Trans. Ultrason. Ferroelectr. Freq. Control 57 2522Google Scholar
[23] 孙宝申, 沈建中 1993 应用声学 3 43Google Scholar
Sun B S, Shen J Z 1993 Appl. Acoust. 3 43Google Scholar
[24] Trots I, Nowicki A, Lewandowski M 2010 WASET 4 136
[25] 康荣宗, 田鹏武, 于宏毅 2014 63 200701Google Scholar
Kang R Z, Tian P W, Yu H Y 2014 Acta Phys. Sin. 63 200701Google Scholar
[26] 刘珍黎, 宋亮华, 白亮, 许凯亮, 他得安 2017 66 154303Google Scholar
Liu Z L, Song L H, Bai L, Xu K L, Ta D A 2017 Acta Phys. Sin. 66 154303Google Scholar
[27] Liu C C 2014 Ph. D. Dissertation (Shanghai: Fudan University) (in Chinese)
[28] Baron C, Talmant M, Laugier P 2007 J. Acoust. Soc. Am. 122 1810Google Scholar
[29] Pithioux M, Lasaygues P, Chabrand P 2002 J. Biomech. 35 961Google Scholar
[30] Qin K H, Yang C, Sun F 2014 IEEE Trans. Ultrason. Ferroelectr. Freq. Control 61 133Google Scholar
Catalog
Metrics
- Abstract views: 9567
- PDF Downloads: 139
- Cited By: 0